Force CT Turbo Flash扫描模式联合迭代重建算法降低辐射剂量:主动脉CTA的比较研究
  • ISSN号:2095-4344
  • 期刊名称:《中国组织工程研究》
  • 时间:0
  • 分类:R816.2[医药卫生—放射医学;医药卫生—临床医学]
  • 作者机构:[1]内蒙古包头市中心医院放疗科,内蒙古包头014040, [2]内蒙古医科大学基础医学院解剖学教研室,内蒙古包头014040, [3]内蒙古医科大学数字医学中心,内蒙古包头014040
  • 相关基金:国家自然科学基金项目(80260269,81560348,81660358)
中文摘要:

目的:利用Force CT的Turbo Flash模式行主动脉全程扫描成像,对照评价其图像质量,旨在评价该模式行主动脉全程扫描联合迭代重建算法降低辐射剂量的可行性.方法:选取内蒙古包头市中心医院2015-08/2016-12行主动脉全程CTA扫描的患者60例,将其分为实验组用Turbo Flash模式扫描(A组,n=30),使用迭代重建算法(ADMIRE)重建图像、对照组用Body Angio模式扫描(B组,n=30),使用滤波反投影算法(FBP)重建图像,并用客观和主观两种方式评价其CT图像,前者包括CNR、SNR及图像噪声,而后者则由两位CT影像诊断副主任医师采用双盲法分别评价.结果:Turbo Flash模式联合迭代重建算法图像的主观评分明显高于Body Angio扫描组,差异具有统计学意义(P〈0.05),而两组图像质量的客观评价比较,差异无统计学意义(P〉0.05).Turbo Flash模式组的扫描时间(1.0±0.2)s,相对于Body Angio模式(4.6±1.4)s明显缩短,差异具有统计些意义(P〈0.05).有效辐射剂量Turbo Flash模式组为(3.7±1.2)mSv,较Body Angio模式组的(6.8±2.8)mSv明显减少,差异具有统计学意义(P〈0.05).结论:Force CT Turbo Flash模式主动脉CTA扫描联合迭代重建算法的图像质量可满足对于主动脉病变的诊断需求,具有扫描时间短、辐射剂量低、运动伪影少等优势,可成为主动脉CTA优先选择的扫描模式推广应用.

英文摘要:

AIM: To investigate the feasibility,image quality and radiation dose for tube-flash protocol of force-CT with ADMIRE of the whole aorta. METHODS: Each group of 30 patients underwent CTA either on force-CT device with tube-flash protocol with ADMIRE( group A) or Body Angio protocol with FBP( group B). Evaluation of its image quality and radiation dose was evaluated objectively and subjectively. Objective evaluation indices included CNR,noise,SNR. Subject image quality of the aorta was independently assessed by two associate chief diagnosticians of CT imaging in a double-blind manner. Scanner-reported dose-length product values were used to estimate effective dose values. Radiation dose evaluation indices included CTDIvol,DLP and ED. RESULTS: The subjective scoring of image quality in the high-pitch examination protocol was obviously higher than that of group B,and the difference was statistically significant(P〈0.05). There was no statistically significant difference on objective evaluation of image quality between the two groups(P〈0.05).The imaging time of group A was( 1.0±0.2) s,which was significantly shorter than that of group B( 4.6±1.4) s,with a statistically significant difference(P〈0.05). Compared with group B( 6.8 ±2.8) mS v,ED of group A( 3.7 ± 1.2) mSv was significantly decreased,and the difference was statistically significant(P〈0.05). CONCLUSION: CT angiography of the whole aorta using tube-flash protocol of force-CT with ADMIRE can meet the need of patients with aortic diseases with shorter scan time,lower radiation dose and less motion artifacts,which has the potential to become a preferred application of aorta CTA.

同期刊论文项目
同项目期刊论文
期刊信息
  • 《中国组织工程研究》
  • 北大核心期刊(2014版)
  • 主管单位:中华人民共和国卫生和计划生育委员会
  • 主办单位:中国康复医学会 《中国组织工程研究》杂志社
  • 主编:唐佩福
  • 地址:沈阳浑南新区10002邮政信箱
  • 邮编:110180
  • 邮箱:crter3377@163.com
  • 电话:024-31416864
  • 国际标准刊号:ISSN:2095-4344
  • 国内统一刊号:ISSN:21-1581/R
  • 邮发代号:8-584
  • 获奖情况:
  • 2001“百种中国杰出学术期刊,卫生部首届医药卫生优秀获奖期刊,北方优秀期刊,辽宁省一级期刊,第三、四届沈阳市优秀期刊一等奖
  • 国内外数据库收录:
  • 美国化学文摘(网络版),荷兰文摘与引文数据库,荷兰医学文摘,中国中国科技核心期刊,中国北大核心期刊(2008版),中国北大核心期刊(2014版)
  • 被引量:16688