不同能量放射性核素在PET/CT及SPECT采集中的相互影响
  • ISSN号:1005-5185
  • 期刊名称:《中国医学影像学杂志》
  • 时间:0
  • 分类:R814.42[医药卫生—影像医学与核医学;医药卫生—放射医学;医药卫生—临床医学]
  • 作者机构:[1]内蒙古医学院附属医院核医学科PET/CT中心,内蒙古呼和浩特010050, [2]内蒙古医学院附属医院心内科,内蒙古呼和浩特010050
  • 相关基金:内蒙古医疗卫生科研计划项目(2010041); 内蒙古科技厅应用技术与开发科研项目(09sfpet)
中文摘要:

目的 利用模型探讨99TcmO4-与18F- 氟脱氧葡萄糖(FDG)在图像采集中的相互影响,讨论同一患者是否可以同日分别采用 PET/CT 及 SPECT 两种仪器行高低两种能量放射性核素显像。材料与方法 利用统一规格、相同容积(500ml)的 5 个塑料瓶(编号 A、B、C、D、E),其内均注入 500ml 蒸馏水制作成模型。A 瓶注入 0.5mCi18F-FDG 行 PET/CT 显像;B 瓶注入 1mCi99TcmO4-行 SPECT 显像;C 瓶内同时注入混合好的0.5mCi18F-FDG 和 1mCi99TcmO4-;D 瓶内先注入 1mCi99TcmO4-后再注入 0.5mCi18F-FDG ;E 瓶先注入0.5mCi18F-FDG 后再注入 1mCi99TcmO4-。然后对 C、D、E 瓶分别行 PET/CT 显像和 SPECT 显像,并判断图像效果。结果 A 瓶 PET/CT 显像和 B 瓶 SPECT显像图像效果为 1 级;C、D 瓶 SPECT 及 PET/CT 显像图像均为 2 级;E 瓶SPECT 显像图像为 3 级,PET/CT 显像图像为 1 级。结论 同时加入18F-FDG和99TcmO4-对 PET/CT 和 SPECT 图像质量均有影响。先加入18F-FDG 后加入99TcmO4-对 PET/CT 图像质量影响小,但 SPECT 图像不能分辨。先加入99TcmO4-后加入18F-FDG 对 PET/CT 和 SPECT 图像质量均有影响。

英文摘要:

Purpose To evaluate interaction between 99 Tc m O 4-and 18 F-FDG in the image acquisition,and to discuss if the patient is eligible to perform radionuclide imaging with high-energy and low-energy by SPECT and PET/CT respectively on the same day.Materials and Methods The cardiac models were made by 5 plastic bottles with uniform specification and the same volume (500 ml).All the bottles were filled in 500 ml water and numbered A,B,C,D and E,respectively.18 F-FDG was added to bottle A and performed PET/CT.99 Tc m O 4-was added to bottle B and performed SPECT.Mixed 0.5mCi 18 F-FDG and 1mCi 99 Tc m O 4-was added to bottle C,and for bottle D,1mCi 99 Tc m O 4-was added first and followed by 0.5mCi 18 F-FDG.For bottle E,0.5mCi 18 F-FDG was added first and followed by 1mCi 99 Tc m O 4-.Then all of C,D,E bottles performed PET/CT and SPECT imaging respectively,and the grades of images were evaluated.Results PET/CT imaging quality were grade Ⅰ for bottle A and SPECT imaging quality were gradeⅠfor bottle B.Both of PET /CT and SPECT imaging quality were grade Ⅱ for bottle C and bottle D.SPECT imaging quality were grade Ⅲ,and PET /CT images quality were grade Ⅰ for bottle E.Conclusion If 99 Tc m O 4-and 18 F-FDG are adding at the same time,both image quality of PET/CT and SPECT will be reduced.If 18 F-FDG is adding first and followed by 99 Tc m O 4-,PET /CT image quality will be slightly reduced,but SPECT images cannot be distingushed.If 99 Tc m O 4-is adding first and followed by 18 F-FDG,both PET /CT and SPECT imaging quality will be reduced.

同期刊论文项目
同项目期刊论文
期刊信息
  • 《中国医学影像学杂志》
  • 中国科技核心期刊
  • 主管单位:中华人民共和国国家卫生和计划生育委员会
  • 主办单位:中国医学影像技术研究会
  • 主编:蔡幼铨
  • 地址:北京复兴路28号
  • 邮编:100853
  • 邮箱:zyyz@j-medicalimaging.com
  • 电话:010-66939381
  • 国际标准刊号:ISSN:1005-5185
  • 国内统一刊号:ISSN:11-3154/R
  • 邮发代号:82-712
  • 获奖情况:
  • 国内外数据库收录:
  • 被引量:18321