树脂粘固剂厚度对CAD/CAM长石质陶瓷断裂性能的影响
  • ISSN号:1671-167X
  • 期刊名称:《北京大学学报:医学版》
  • 时间:0
  • 分类:R783.1[医药卫生—口腔医学;医药卫生—临床医学]
  • 作者机构:[1]北京大学口腔医学院·口腔医院牙体牙髓科,北京100081, [2]北京市崇文口腔医院口腔内科,北京100062
  • 相关基金:卫生公益性行业科研专项(201002017)资助
中文摘要:

目的:测试树脂粘固剂不同厚度对CAD/CAM陶瓷断裂韧性值的影响,以期对临床医生提供指导和参考。方法:分别制作12 mm×14 mm×2 mm的CAD/CAM陶瓷和树脂片,用树脂粘固剂进行粘接,分为3组:第1组为空白对照,瓷-树脂片间无粘固剂;第2组的瓷-树脂片间粘固剂厚度约为50μm,第3组的瓷-树脂片间粘固剂厚度100μm。用维氏硬度仪测试硬度和断裂韧性,并进行Weibull分析。结果:3组样本的硬度差异无统计学意义,断裂韧性分别为(1.02±0.11)MN/m3/2,(0.99±0.10)MN/m3/2和(0.97±0.14)MN/m3/2,第3组低于第1组(P〈0.05),第2组断裂韧性Weibull分布模数最高,第1~3组的特征断裂韧性值分别为1.08、1.04和1.03MN/m3/2。结论:树脂粘固剂厚度能影响陶瓷的断裂韧性值,提示临床操作应该控制粘固剂的厚度,避免过厚。

英文摘要:

Objective:Chair-side CAD/CAM ceramic restorations are esthetic with a time-saving procedure.However,ceramic restorations often failed due to fracture.This study was to investigate the effect of cement thickness on fracture toughness and hardness of CAD/CAM ceramic.Methods: Machinable ceramic blocks for CAD/CAM(Sirona CEREC Blocs) were cut into 9 slices(12 mm×14 mm×2 mm) using a diamond saw(Leica SP 1600) and polished carefully with sandpaper / diamond pastes to yield a mirror-like surface.LIGHT-CORETM was used to prepare resin slices with dimension of 12 mm×14 mm×2 mm.The polished ceramic slices were then randomly divided into 3 groups and cemented to resin slices using Multilink Automix: Group 1,no cement;Group 2,the thickness of cement was 50 μm;Group 3,the thickness of cement was 100 μm.Hardness(H) was determined by Vickers indentation technique using micro-hardness tester(HMV-2T Shimadzu) with a load of 19.6 N for 15 s.The fracture toughness(KIC) was calculated by KIC = 0.016(E/H)0.5(P/c1.5),where P= applied load,c= crack length,E=elastic modulus,H=(0.47 P/a2),a= half diagonal of the indentation.Results: The fracture toughness for Group 1(no cement) was(1.02±0.11) MN/m3/2,for Group 2 and Group 3 was(0.99±0.10) MN/m3/2 and(0.97±0.14) MN/m3/2.Group 3 showed lower fracture toughness than group 1(P0.05).The Weibull parameter m of group 2 was the highest and K0 of three groups were 1.08 MN/m3/2,1.04 MN/m3/2 and 1.03 MN/m3/2 respectively.The hardness for Group 1 was(6.68±0.73) GPa,group 2 [(6.85±0.64) GPa],group 3 [(6.81±0.98) GPa].There was no significant deference in hardness among the 3 groups(P0.05).Conclusion: Thickness of cementation may affect fracture toughness of machinable ceramic,therefore,a thinner film of cementation is recommended for chair-side CAD/CAM ceramic restorations.

同期刊论文项目
期刊论文 45
同项目期刊论文
期刊信息
  • 《北京大学学报:医学版》
  • 北大核心期刊(2011版)
  • 主管单位:中华人民共和国教育部
  • 主办单位:北京大学
  • 主编:韩启德
  • 地址:北京海淀区学院路38号
  • 邮编:100191
  • 邮箱:
  • 电话:010-82801551
  • 国际标准刊号:ISSN:1671-167X
  • 国内统一刊号:ISSN:11-4691/R
  • 邮发代号:2-489
  • 获奖情况:
  • 1992年全国优秀科技期刊评比三等奖,1996年第二届全国优秀科技期刊评比二等奖,2001年入选中国期刊方阵,被评为“双百”期刊,2007年获得第六届百种中国杰出学术期刊奖
  • 国内外数据库收录:
  • 美国化学文摘(网络版),波兰哥白尼索引,荷兰文摘与引文数据库,美国生物医学检索系统,日本日本科学技术振兴机构数据库,中国中国科技核心期刊,中国北大核心期刊(2004版),中国北大核心期刊(2008版),中国北大核心期刊(2011版),中国北大核心期刊(2014版)
  • 被引量:16532