位置:成果数据库 > 期刊 > 期刊详情页
Uncertainty of an automatic system for counting alpha tracks on CR-39
  • ISSN号:1004-714X
  • 期刊名称:《中国辐射卫生》
  • 时间:0
  • 分类:TL[核科学技术]
  • 作者机构:[1]Institute of Radiation Medicine, Fudan University,Shanghai 200032, China
  • 相关基金:This work was supported by the National Natural Science Foundation of China (No. 11375048).
中文摘要:

For accurate counting of alpha tracks on the polyallyl diglycol carbonate of CR-39-type track detectors,the size distributions of both artifact tracks and alpha tracks were investigated with an automatic counting system. At the same temperature and etchant concentration, the numbers and sizes of alpha tracks changed significantly with the etching time, and the artifact track changes were smaller. At the etching time of 5 h, the sizes of alpha tracks were evidently larger than those of the artifact tracks, and the deviation of its size distribution was much smaller than those of longer etching time. Based on the size distribution of alpha tracks etched for 5 h, the overlap effect and uncertainty of overlap correction were studied by the Monte Carlo simulations for different track densities. It was found that the counting uncertainty of the system could be less than 6% in a density range of 10–160 tracks mm-2 after taking the overlap correction into account.

英文摘要:

For accurate counting of alpha tracks on the polyallyl diglycol carbonate of CR-39-type track detectors, the size distributions of both artifact tracks and alpha tracks were investigated with an automatic counting system. At the same temperature and etchant concentration, the numbers and sizes of alpha tracks changed significantly with the etching time, and the artifact track changes were smaller. At the etching time of 5?h, the sizes of alpha tracks were evidently larger than those of the artifact tracks, and the deviation of its size distribution was much smaller than those of longer etching time. Based on the size distribution of alpha tracks etched for 5?h, the overlap effect and uncertainty of overlap correction were studied by the Monte Carlo simulations for different track densities. It was found that the counting uncertainty of the system could be less than 6% in a density range of 10-160 tracks?mm?2 after taking the overlap correction into account.

同期刊论文项目
同项目期刊论文
期刊信息
  • 《中国辐射卫生》
  • 北大核心期刊(2004版)
  • 主管单位:中华人民共和国卫生和计划生育委员会
  • 主办单位:中华预防医学会 山东省医学科学院放射医学研究所
  • 主编:
  • 地址:济南市经十路18877号
  • 邮编:250062
  • 邮箱:REDI@chinajournal.net.cn
  • 电话:0531-82919955
  • 国际标准刊号:ISSN:1004-714X
  • 国内统一刊号:ISSN:37-1206/R
  • 邮发代号:24-94
  • 获奖情况:
  • 中华预防医学会系列杂志优秀期刊
  • 国内外数据库收录:
  • 中国中国科技核心期刊,中国北大核心期刊(2004版)
  • 被引量:6585