位置:成果数据库 > 期刊 > 期刊详情页
红细胞比容国家一级标准物质的研制
  • ISSN号:1671-2587
  • 期刊名称:《临床输血与检验》
  • 时间:0
  • 分类:R331.141[医药卫生—人体生理学;医药卫生—基础医学]
  • 作者机构:[1]卫生部北京医院、卫生部临床检验中心,100730
  • 相关基金:本课题受国家自然科学基金(30972818)和国家科技基础性工作专项(No.2013FY113800)资助
中文摘要:

目的 研制红细胞比容国家一级标准物质,作为红细胞比容检测结果溯源的标准.方法 新鲜血经稳定化处理、调整浓度后分装成待评价的标准物质;参照ISO Guide 35(2006版)和《一级标准物质技术规范》的要求进行均匀性和稳定性评价,使用直接溯源至参考方法的标准仪器对标准物质进行定值;对标准物质的均匀性、稳定性和定值过程引入的不确定度进行评定,计算合成标准不确定度和扩展不确定度;将一级标准物质与二级标准物质的不确定度进行比较.结果 标准物质的均匀性评价方差分析结果显示P>0.05,表明标准物质是均匀的;瓶间均匀性的标准不确定度ubb为0.22%.12周内稳定性评价趋势分析结果显示P>0.05,表明该期限内标准物质是稳定的;长期稳定性的标准不确定度ults为0.27%.定值结果的标准不确定度uchar为0.59%;合成并扩展不确定度后,定值结果的表达式为(38.7±1.4)%.一级标准物质的相对不确定度(3.6%)小于国家二级标准物质(4.8%).结论 标准物质的均匀性和稳定性良好,定值方法可靠.

英文摘要:

Objective To study and prepare hematocrit primary reference material for traceability.Methods Hematocrit primary reference material was prepared by stabilizing the blood cells,adjusting the concentration,and subpackaging the blood.The candidates were evaluated according to ISO Guide 35 and "Technical Norm of Primary Reference Material",including homogeneity and stability studies,and characterization by the standard analyzer which is traceable to reference method directly.The uncertainty from homogeneity,stability,and characterization was evaluated.The combined uncertainty and extended uncertainty were calculated.The uncertainty between prepared reference material and the secondary reference material were compared.Results The results of one-way analysis of variance of hematocrit showed that the reference material was homogeneous (P〉0.05).The bottle to bottle homogeneity uncertainty (ubb) was 0.22%.The reference material was stable in 12 weeks by trend analysis (P〉0.05).The stability uncertainty (ults) was 0.27%.The characterization uncertainty (uchar) was 0.59%.The assigned value was found to be (38.7± 1.4)% through calculation of the combined uncertainty and extended uncertainty.The uncertainty of the primary reference material(3.6%) is smaller than that of the secondary reference material(4.8%).Conclusion The candidate reference material had promising homogeneity and stability,and the method of characterization of the candidates was accurate and reliable.

同期刊论文项目
同项目期刊论文
期刊信息
  • 《临床输血与检验》
  • 中国科技核心期刊
  • 主管单位:安徽省卫生厅
  • 主办单位:安徽省输血协会 安徽省立医院
  • 主编:权循珍
  • 地址:合肥市庐江路17号安徽省立医院内
  • 邮编:230001
  • 邮箱:cbtls01@163.com
  • 电话:0551-62282341
  • 国际标准刊号:ISSN:1671-2587
  • 国内统一刊号:ISSN:34-1239/R
  • 邮发代号:26-186
  • 获奖情况:
  • 国内外数据库收录:
  • 美国化学文摘(网络版),中国中国科技核心期刊
  • 被引量:7822