血管位置对HIFU焦域温度场的影响
  • 期刊名称:国际生物医学工程杂志
  • 时间:0
  • 页码:16-19
  • 语言:中文
  • 分类:R318[医药卫生—生物医学工程;医药卫生—基础医学] Q681[生物学—生物物理学]
  • 作者机构:[1]天津医科大学生物医学工程学院,300070, [2]天津医科大学肿瘤医院,300040
  • 相关基金:基金项目:国家自然科学基金资助项目(30870718);天津市应用基础及前沿技术研究计划重点项目(08JCZDJC25000)
  • 相关项目:血流、组织声学特性对HIFU焦域温度分布影响的研究
中文摘要:

目的数值仿真研究不同位置的血管对高强度聚焦超声(HIFU)可治疗焦域的影响,为HIFU治疗安全性和可靠性提供理论依据。方法应用超声非线性传播方程、Pennes生物热传导方程以及FDTD法数值仿真温度场分布。结果考虑血流灌注与否,HIFU焦域温度场分布相一致。血管中心位于声轴时,焦域中心温升非常低;当血管偏离声轴小于3.0mm时,可治疗焦区不再关于声轴对称,焦域中心温升没有明显变化;当血管偏离声轴大于3.0mm时,血管对可治疗区域的温度场分布基本无影响。结论脂肪组织内血流灌注对HIFU焦域温度场的影响可以不予考虑;直径2mm血管位于可治疗区域内,对血管内及周边组织温升影响很大,可能导致血管周围组织不完全致死;当直径2mm血管距声轴小于3mm对可治疗区域仍有影响,但大于3.0mm时,血管对可治疗区域的影响可以不予考虑。

英文摘要:

Objective To improve the safety and reliability in HIFU therapy, the effects of the vessel locations on the high intensity focused ultrasound temperature field were investigated. Methods The temperature distribution was obtained by combining the nonlinear acoustic propagation formula and the Pennes bio-beat transfer equation, as well as the finite difference time domain method. Results The HIFU temperature fields were almost unchangeable whether the vessels had perfusion or not. However, there was a significant reduction in the center temperature rise, while the vessel was located in the acoustic axis. If the distance between the vessel wall and the acoustic axis was less than 3.0mm, the therapeutic focal region was no longer symmetric and the change in the peak temperature rise was negligible. Neither did the change in the temperature field if the distance was larger than 3.0 mm. Conclusion The effects of vessel perfusion on the HIFU temperature field may be ignored, When the vessel is located within the therapeutic region, the one with 2 mm diameter has a significant effect on the temperature rise of the endovaseular and the surrounding tissues, leading to incomplete infarct of the tissues around the vessel. When the distance is less than 3.0mm, the vessel with 2 mm diameter might affect the distribution of temperature field. While the effect is negligible, if the distance is larger than 3.0 ram.

同期刊论文项目
同项目期刊论文