A gamma spectrum of a Pu-C source is measured using a p-type HPGe detector, whose three peaks (full energy, single-escape and double-escape peak) can be used as a calibration source for the beam energy measurement system of BEPCII. The effect of fast neutron damage on the energy resolution of the HPGe detector is studied, which indicates that the energy resolution begins to deteriorate when the detector is subject to 2×107 n/cm2 fast neutrons. The neutron damage mechanism and detector repair methods are reviewed. The Monte Carlo simulation technique is utilized to study the shielding of the HPGe detector from the fast neutron radiation damage, which is of great significance for the future commissioning of the beam energy measurement system.
A gamma spectrum of a Pu-C source is measured using a p-type HPGe detector, whose three peaks (full energy, single-escape and double-escape peak) can be used as a calibration source for the beam energy measurement system of BEPCII. The effect of fast neutron damage on the energy resolution of the HPGe detector is studied, which indicates that the energy resolution begins to deteriorate when the detector is subject to 2×107 n/cm^2 fast neutrons. The neutron damage mechanism and detector repair methods are reviewed. The Monte Carlo simulation technique is utilized to study the shielding of the HPGe detector from the fast neutron radiation damage, which is of great significance for the future commissioning of the beam energy measurement system.