分子印迹电化学传感器选择性识别及电催化检测多巴胺
  • ISSN号:0567-7351
  • 期刊名称:化学学报
  • 时间:2013.6.15
  • 页码:951-956
  • 分类:O614.121[理学—无机化学;理学—化学]
  • 作者机构:[1]桂林理工大学化学与生物工程学院,桂林541004
  • 相关基金:项目受国家自然科学基金(N0.21165007)和广西自然科学基金(No.2012GxNsFAA053032)资助.
  • 相关项目:基于放大效应的磁纳米粒子分子印迹电化学传感器研究
中文摘要:

以多巴胺为模板分子、溴酚蓝为单体, 在玻碳电极表面制得分子印迹膜. 洗脱后重新吸附多巴胺分子, 同时利用聚溴酚蓝膜对多巴胺氧化的电催化作用, 对多巴胺进行定量分析. 多巴胺在0~1.2×10-6 mol/L浓度范围内与电流变化值有线性关系. 由于印迹膜的催化作用, 检出限达1.62×10-10 mol/L. 该传感器可用于选择性测定多巴胺, 抗坏血酸和尿酸等共存物不干扰测定. 利用该传感器直接测定了血清中多巴胺含量, 回收率在95.2%~103.0%之间.

英文摘要:

A novel strategy for improving the sensitivity of molecular imprinted electrochemical sensor was proposed and dopamine (DA) was selected as the template molecular in this assay. The electroactive membrane of poly-bromophenol blue (BB) which was polymerized on the electrode surface acted as molecular imprinted membrane of DA. To prepare poly-BB-DA molecularly imprinted polymer (MIP), CV scans were performed for 30 cycles in the potential range between -1.0 and 1.8 V at 50 mV/s in an acetate buffer solution (pH=4.0) containing 1.0×10-3 mol/L DA and 3.0×10-3 mol/L BB at 25 ℃. The MIP electrode was washed by methanol (50% in volume) for 12 min to remove template molecules. Differential pulse voltammetry (DPV) was performed after the rebinding reaction of the DA and the MIP membrane in DA sample solutions for 6 min. With the increasing of the DA concentration, the binding sites in the membrane taken by DA molecules also increased, so was the peak current in the DPV analysis. The sensitivity was improved significantly due to the signal amplifying effect produced by the catalytic effect of electro oxidation of dopamine on BB membrane. The experimental conditions were also optimized. Electrochemical measurements for the MIP membrane characterization were carried out in the supporting electrolyte of 0.01 mol/L K3[Fe(CN)6] solution containing 0.5 mol/L KCl. CV was performed from -0.2 to 0.6 V at a scan rate of 100 mV/s. DPV was performed in the supporting electrolyte of 0.1 mol/L PBS (pH 7.4) over a potential range of -0.1 to 0.6 V, with the pulse amplitude of 50 mV and the scan rate of 50 mV/s. All measurements were carried out at room temperature (25 ℃). DA was determined by DPV, and there was a linear relationship between oxidation currents and DA concentrations in the range of 0~1.2×10-6 mol/L with a detection limit of 1.62×10-10 mol/L. This is one of the most sensitive sensors for DA determination due to the catalytic oxidation of DA on electroactive MIP.

同期刊论文项目
同项目期刊论文
期刊信息
  • 《化学学报》
  • 北大核心期刊(2014版)
  • 主管单位:中国科学院
  • 主办单位:中国化学会 中国科学院上海有机化学研究所
  • 主编:周其林
  • 地址:上海市零陵路345号
  • 邮编:200032
  • 邮箱:hxxb@sioc.ac.cn
  • 电话:021-54925085
  • 国际标准刊号:ISSN:0567-7351
  • 国内统一刊号:ISSN:31-1320/O6
  • 邮发代号:4-209
  • 获奖情况:
  • 首届国家期刊奖,第二届国家期刊奖提名奖,中国期刊方阵“双高期刊”
  • 国内外数据库收录:
  • 俄罗斯文摘杂志,美国化学文摘(网络版),荷兰文摘与引文数据库,美国科学引文索引(扩展库),日本日本科学技术振兴机构数据库,中国中国科技核心期刊,中国北大核心期刊(2004版),中国北大核心期刊(2008版),中国北大核心期刊(2011版),中国北大核心期刊(2014版),英国英国皇家化学学会文摘,中国北大核心期刊(2000版)
  • 被引量:28694